NewLife® Intensity 10 Oxygen Concentrator Service Manual AirSep Corp

Página 2 - 08/11

MN134-1 rev A 08/11 5 1.4 Functional Specifications Oxygen Concentration: 2-9 LPM: 92% +3.5/-3% 10 LPM: 90% +5.5/-3% (Based on 70

Página 3 - Table of Contents

MN134-1 rev A 08/11 6 2.0 Operational Check and Oxygen Concentration Test 2.1 Description of Operation Air is drawn into the NewLife Intensi

Página 4 - Section 5.0

MN134-1 rev A 08/11 7an intermittent alarm if the high or low pressure indicators are activated or if the optional oxygen monitor detects lower than

Página 5 - Appendix

MN134-1 rev A 08/11 8 3.0 Patient Instructions 3.1 Instructions It is important that patients thoroughly understand how to operate the AirSep Ne

Página 6 - 1.0 Introduction

MN134-1 rev A 08/11 9 4.0 Provider Instructions 4.1 Instructions To ensure that the unit’s output of oxygen is within specification, you must pe

Página 7 -

MN134-1 rev A 08/11 104.1.5 Recording Maintenance As the Equipment Provider, you can record all routine maintenance and repairs performed on the Ne

Página 8

MN134-1 rev A 08/11 115.0 Main Components 5.1 Components The design of the AirSep NewLife Intensity 10 Oxygen Concentrator allows for easy access

Página 9

MN134-1 rev A 08/11 12 5.3 Compressor The compressor is the “pump” within the oxygen concentrator that pushes the room air into the bottom of the s

Página 10

MN134-1 rev A 08/11 135.3.2 Capacitor Replacement The capacitor starts the compressor. If the compressor cannot start, the capacitor may be defec

Página 11 - 2.3 Alarm System

MN134-1 rev A 08/11 14b. Feed/waste coil = 850ohms +/- 10% 2. 220-240VAC units a. EQ coil = 4,300ohms +/- 10% b. Feed/waste coil = 3,100ohms +/

Página 12 - 2.3.1 Start Up/Battery Test

MN134-1 rev A 08/11 ii Expedited Product Warranty Check service is always at your fingertips with AirSep: Type: http://www.airsep.com/medica

Página 13 - 3.0 Patient Instructions

MN134-1 rev A 08/11 15 If replacement is necessary, you must replace both sieve beds at the same time. Sieve Bed Removal To remove sieve beds,

Página 14 - 4.0 Provider Instructions

MN134-1 rev A 08/11 16A green light that fails to illuminate indicates a disconnected or faulty solenoid coil or an electrical malfunction in that v

Página 15 - Infection Control

MN134-1 rev A 08/11 174. Remove the 2nd stage intake resonator, which is held in place by Velcro. 5. Remove the securing nut that holds the produc

Página 16 - 5.0 Main Components

MN134-1 rev A 08/11 18 5.9.2 Product Regulator Cleaning or Rebuilding Clean or rebuild the product regulator if the regulator cannot be adjusted fo

Página 17 - 5.3 Compressor

MN134-1 rev A 08/11 19 I/0 Power Switch Installation Follow the removal procedure for the I/0 power switch in reverse order to install a new power

Página 18 - 5.4 Solenoid Valves

MN134-1 rev A 08/11 203. Open the plastic twist clamps. 4. Cut the tie-wraps (located near the power cord push-on connectors). 5. Disconnect the

Página 19 - 5.5 Sieve Beds

MN134-1 rev A 08/11 216.0 Troubleshooting 6.1 Operating Pressure Test Testing the operating pressure is a useful diagnostic tool when a concentr

Página 20 - 5.7 Circuit Board

MN134-1 rev A 08/11 226.2 General Troubleshooting Before reviewing the troubleshooting chart, the following steps may be useful to isolate any malf

Página 21 - ESD safety procedures

MN134-1 rev A 08/11 236.3 Troubleshooting Chart Problem Probable Cause Solution Unit does not run. Constant audio alarm with I/0 power switch in

Página 22

MN134-1 rev A 08/11 24 Problem Probable Cause Solution Compressor runs with intermittent high pressure alarm and low oxygen concentration. Restr

Página 23 - I/0 Power Switch Removal

MN134-1 rev A 08/11 iiiTable of Contents Section 1.0 Introduction 1.1 Equipment Provider Responsibility 1 1.2 Important Notice a

Página 24 - 5.15 Power Cord Replacement

MN134-1 rev A 08/11 25 Problem Probable Cause Solution Cabinet fan does not turn. Faulty electrical connections. Faulty cabinet fan. Check elec

Página 25 - (In Equipped Units)

MN134-1 rev A 08/11 26 6.4 Tool Kit and Pressure Test Gauge The tools needed for you to properly service the NewLife Intensity 10 unit are listed b

Página 26 - 6.0 Troubleshooting

APPENDIX  Exploded drawings Wiring Diagram – Intensity 10 120V A Wiring Diagram – Intensity 10 220/240V B Control Panel Assembly

Página 27 - 6.2 General Troubleshooting

MN134-1 rev A 08/11 A WIRING DIAGRAM INTENSITY 10 120V

Página 28 - 6.3 Troubleshooting Chart

MN134-1 rev A 08/11 B WIRING DIAGRAM INTENSITY 10 220/240V

Página 29 - /0 power switch

MN134-1 rev A 08/11 CITEM PART NUMBER DESCRIPTION ITEMPART NUMBER DESCRIPTION 1 BK016-1 BRACKET, FLOWMETER, DUAL 13 FA002-1 FASTENER, TIE WRAP

Página 30

MN134-1 rev A 08/11 DITEM PART NUMBER DESCRIPTION ITEMPART NUMBER DESCRIPTION 1 BT001-2 BATTERY, HOLDER 15 SC002-1 SCREW, TEMP SWITCH 2 BT002-1 BA

Página 31

MN134-1 rev A 08/11 EITEM PART NUMBER DESCRIPTION ITEMPART NUMBER DESCRIPTION 1 BE186-1R BEDS 6A CD011-2 CORD, POWER, US-STYLE 2 CA034-1 CABINET

Página 32 - APPENDIX

MN134-1 rev A 08/11 FITEM PART NUMBER DESCRIPTION 1 RG088-1 REGULATOR (ONLY) 2 TA132-1 TANK, MIXING, ASSY PRODUCT TANK/REGULATOR ASSEMBLY

Página 33 - INTENSITY 10 120V

MN134-1 rev A 08/11 G ITEM PART NUMBER DESCRIPTION ITEMPART NUMBER DESCRIPTION 1 VA290-3 VALVE, BLOCK (ONLY) 3B VA117-1 VALVE, COIL, 220V 2 V

Página 34 - INTENSITY 10 220/240V

MN134-1 rev A 08/11 ivSection 5.0 Main Components 5.1 Components 11 5.2 Cabinet Removal 11 5.2.1 Removing Side Panel(s) 11 5.2.2 Rem

Página 35

MN134-1 rev A 08/11 HITEM PART NUMBER DESCRIPTION ITEMPART NUMBER DESCRIPTION 1 CL007-1 CLAMP, HEAT EXCHANGER 6 F0104-1 FITTING, ELBOW, HEAT E

Página 36 - MAIN STRUCTURE ASSEMBLY

MN134-1 rev A 08/11 IITEM PART NUMBER DESCRIPTION ITEMPART NUMBER DESCRIPTION 1 BE002-4 BED, TUBE 9 F0001-2 FITTING, ELBOW, TOP CAP 2 BE005-2

Página 37 - BASE AND CABINET COMPONENTS

MN134-1 rev A 08/11 vSection 6.0 Troubleshooting 6.1 Operating Pressure Test 21 6.1.1 High Operating Pressure 21 6.1.2 L

Página 38 - DESCRIPTION

MN134-1 rev A 08/11 11.0 Introduction 1.1 Equipment Provider Responsibility All Equipment Providers of the NewLife® Intensity 10 Oxygen Concentra

Página 39 - VALVE BLOCK ASSEMBLY

MN134-1 rev A 08/11 21.2 Important Notice and Symbol Explanations As you read the manual, pay special attention to the WARNING, CAUTION, and NOTE m

Página 40 - COMPRESSOR ASSEMBLY

MN134-1 rev A 08/11 3Method for disposing of waste: All waste from the NewLife Intensity 10 (patient circuit, etc.) must be disposed of using approp

Página 41 - SIEVE BED ASSEMBLY

MN134-1 rev A 08/11 41.3 NewLife Intensity 10 Product Warranty AirSep Corporation warrants the NewLife Intensity 10 Oxygen Concentrator to be free f